ZL-2428A Vibrating Test Sieve

Products

ZL-2428A Vibrating Test Sieve

Inspection and analysis sieves are widely used in laboratories, quality inspection rooms and other inspection departments to determine the particle size distribution of granules and powder materials, and the determination and analysis of product impurity content and liquid solid content. The product has the characteristics of compact structure, low noise, high efficiency and accurate precision. It adopts a high-quality new type of power generator to generate multiple high-frequency vibrations to efficiently screen and measure materials. At the same time, different working hours can be realized through the electronic delay device according to different materials, so as to minimize the inspection error, ensure the consistency of sample analysis data, and make accurate judgments on product quality.  

  1. Round handle 2. Fastening handle 3. Press head 4. Upright rod 5. Screening tool 6. Nut 7. Tray 8. Vibrating body 9. Spring
  2. Spring seat 11. Bottom cylinder 12. Foot 13. Shell 14. Vibration motor 15. Relay 16. Switch 17. Cover

Instructions

  1. Determine the standard sieve to be selected according to the inspected material and the corresponding standard. 
  2. Stack the standard sieves on the tray seat in order from large to small, from bottom to top, and position the standard sieves by grooves or positioning screws. 
  3. Put the material to be tested into the standard sieve at the top, (or cover the upper cover of the standard sieve), then press the standard sieve with the sieve head set on the wire column, and screw it tightly.

The nut on the spool to compress the standard screen. (Note: Both sides should be the same force, and then use the locking screw to lock.)

  1. Put the timer switch in the corresponding required position, (pay attention to reading the timer description, the time set by different settings is different) and then turn on the power switch,

The inspection sieve starts working. 

  1. After the standard inspection sieve stops working, unscrew the nut on the wire column, remove the sieve head, and carefully remove the standard sieve. 
  2. Cut off the power. 

Precautions

  1. Standard analytical sieves are required to be placed horizontally. 
  2. Confirm that the power supply meets the requirements and ensure that it is grounded. 
  3. The vibration part cannot be in contact with other objects. 

Fourth, the standard screen must be positioned by grooves or positioning screws. 

troubleshooting

State Inspection Items Countermeasures

Poor motor operation Confirm the power supply Confirm the power supply, open the card

Poor contact Check wiring

Single-phase operation Replace the motor

Abnormal sound Nut is not pressed Nut is pressed

Standard screen not positioned Standard screen positioned

Broken spring Replace spring

Contact with other objects Separate from other objects

Loosen the fixing motor screw Lock the screw

Free Quote & Information Request